Thank you for choosing this equipment. We sincerely wish that you enjoy your electrophoresis
power supply Nahita. We highly recommend that you look after the equipment according to what
is stated in this manual.
Nahita develops its products following the CE guidelines taking into account the ergonomics and
safety of the user. The quality of the manufacturing materials and the correct use will allow you
to enjoy this equipment for many years.
The misuse of this equipment may cause accidents, electric shocks, shorts, fires, injuries, etc.
Check "Safety" to read safety guidelines.
Safety
Caution/Warning
Electrophoresis power supplies use high output voltages that are electrically
isolated from earth ground through a protective impedance to minimize the risk of
electrical shock to the user. The following guidelines should be observed and followed
when using a power supply.
Electrophoresis power supplies have passed tests for operation at temperatures
between 0° and 40°C, with non-condensing relative humidity between 0 and 95%.
Operating the power supply outside these conditions is not recommended by Nahita
and will void the warranty.
1.
To ensure adequate cooling of the power supply, be sure that there is at least 6 cm clearance
around the power supply. Do not block the fan vents at the rear of the unit.
2.
Always connect the power supply to a 3-prong, grounded AC outlet, using the3-prong AC
power cord provided with the power supply.
3.
Do not operate the power supply in extreme humidity (95%) or where condensation can
short the internal electrical circuits of the power supply.
4.
When taking the power supply into a cold room, the unit can be operated immediately.
However, when removing the power supply from the cold room, let the unit equilibrate to room
temperature for a minimum of 2 hours before using it.
5.
Never connect a high voltage output lead to earth ground. This defeats the floating electrical
isolation of the power supply and exposes the user to potentially lethal high voltages.
Important
This instrument is intended for laboratory use only. This product conforms to the class A standards
for Electromagnetic Emissions, intended for laboratory equipment applications. It is possible that