The cursor will flash at "U=□□□V". Set the value of constant voltage by pressing "▲" or "▼"
2.2
button. The voltage range can be set from 10 to 600V.
The cursor will flash at "I = □□□mA" when you press "ENT" button. Set the value of constant
2.3
current by pressing "▲" or "▼" button. The current range can be set from 1 to 500mA.
The cursor will flash at "P = □□□W" when you press "ENT" button. Set the value of constant
2.4
power by pressing "▲" or "▼" button. The power range can be set from 1 to 300W.
The cursor will return and flash at "U = □□□V" when you press "ENT" button. Set the value
2.5
of constant voltage again by pressing "▲" or "▼" button.
2.6
Power supply in Setting State now:
Press "RUN/STOP" button, it will enter into Operating State.
2.7
Power supply in Operating State now:
Press "RUN/STOP" button, it will stop output and return to Setting State.
In Setting State, continue to press "EDIT" button, it will enter into Editing State and display:
2.8
"SAVE [□]" → "LOAD [□]" → "T= □□:□□" → "QUIT" → "SAVE [□]"...
At "SAVE □", press "▲"or "▼" button to select the number of memory program. The range
2.9
of numbers is from 1 to 9. Then press "ENT" button to save current method into this
program.
EDIT[0]
U= 100V
I= 500mA
P= 300W
EDIT[ 0 ]
U= 100V
I= 500mA
P= 300W
RUN!
00:01
100
V
350
mA
35
W
0
V
0
mA
0
W
14