3. Store the reference power value: Press the [Ref] key for several seconds
until "HELD" appears on the LCD screen, Prolite-23 will set the current
measured dBm as reference value. The display should read 0.00dB, as
shown in figure 11.
4. Measure Connector Insertion Loss: Place between both instruments the
connector under test to measure so as shows figure 12. The power meter will
read the lost ones of insertion of the connector.
06-2008
Figure 11.- Save reference power value
Page 13