Basic Specifications

Specification*
Frequency
Waveform
Voltage Output Range
Voltage Pulse Width
Current Pulse Range
Current Pulse Width
Maximum Output Charge 25 µC into a load of 500 ohms
Display
Power Source
Standard: ANSI/AAMI
NS4-1985
Dimensions
Weight
Electrical Protection
Rating
*Electrical specifications are +/- 10% into a 500 ohm load unless otherwise noted.
ms = milliseconds
µC = microcoulombs
Output Waveform
The signal is a spiked shape, monophasic, asymmetrical, DC output.
Output
18
®
Hand System
BioniCare Hand System
100Hz ± 5 Hz, fixed digital
Monophasic spike-shaped pulse
0-15 volts peak - digital
1.8 ms @ 10% pt. of peak
0.64 ms @ 50% pt. of peak
0-30 mA peak @ 500 ohms resistive load
1.8 ms @ 10% pt. of peak @ 500 ohms resistive load
0.64 ms @ 50% pt. of peak @ 500 ohms resistive
load
LCD
3.7V Li-Ion battery
Compliant
3.56" (90.5mm) X 2.13" (54.0mm) X 0.59" (15.0mm)
0.152lb or 2.45oz or 69g (With battery)
0.110lb or 1.75oz or 50g (Without battery)
Type BF Equipment, Battery Operated
LCD = liquid crystal display
mA = milliamperes
Width at 10% Peak = 1.8msec
Output Waveform
Figure 5
Width at 50% Peak = 0.64msec

Publicidad