Table of Contents
1 Expanded manual version
2 Specifications
on page 3
3 General information
on page 5
4 Installation
on page 8
5 Analyzer startup
on page 17
Section 1 Expanded manual version
For additional information, refer to the expanded version of this manual, which is available on the
manufacturer's website.
Section 2 Specifications
Specifications are subject to change without notice.
Analyzer
Specification
Details
Dimensions
748 x 250 x 236 mm (29.4 x 9.8 x 9.3 in.)
Weight
7 kg (15.4 lb)
Sample flow rate
5—20 liters/hour
Ambient temperature 0—60 °C (32—140 °F)
Relative humidity
10—90%
Temperature sensor
Pt100
Accuracy
± 1% of displayed value; temperature < ± 0.2 °C
Accuracy on conductivity measurement: ± 2%; Maximum difference calculated-theoretical
Calculated pH
value: 0.1 pH
NH
Display range
NaOH; 7 < pH < 10.7; 2.5 μS/cm < C1 < 125 μS/cm; C2 < 100 μS/cm
Display resolution
Conductivity/resistivity: automatic point drift (minimum resolution 0.001 μS/cm) < 0.1 °C
Polyethylene or PTFE or FEP; 0.2 to 6 bars (3 to 90 psi); 5 to 50 °C (40 to 120 °F); Input:
Sample tubing
6-mm (standard) or 1/4-inch (with adapter); Output: 12-mm or 1/2-inch
Certifications
EN 61326-1: 2006; EN 61010-1: 2010
Sensor
Specification
Sensor body material
Conductivity electrodes, internal and external
Cell constant K
Conductivity range
Maximum pressure
Maximum temperature
Accuracy
on page 3
; 7 < pH < 10; 2.8 μS/cm < C1 < 28 μS/cm; C2 < 0.5 μS/cm
3
Details
Black PSU
Stainless steel 316L
0.01 (cm
0.01—200 μS.cm
10 bar
125 °C (257 °F)
< 2%
6 User interface and navigation
7 Operation
on page 18
8 Maintenance
on page 22
9 Troubleshooting
on page 24
-1
)
-1
; Resistivity range: 5k Ω.cm—100 MΩ.cm
on page 17
English 3