Maintenance; Battery Replacement; Fuse Replacement - Promax PE-455 Manual De Instrucciones

Medidor de aislamiento
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

USER'S MANUAL. PE-455

5. MAINTENANCE

5.1 Battery replacement

ATTENTION:
Disconnect all the test leads before opening the cover to change the
batteries.
1.
Remove the battery cover and the batteries.
2.
Replace with six 1.5 V R6 or LR6 batteries, taking care to observe correct
polarity.
3.
Replace the battery cover.

5.2 Fuse replacement

ATTENTION:
Disconnect all the test leads before opening the cover to change the
fuse.
The fuse is located in the enclosure. To replace the fuse, take the 4 screws out
from the bottom case, then remove and replace the fuse. Make sure to place the fuse
protection cover. (Small rubberised fuse cover). Only replace with the same
specification fuse.
AVOIDING THESE DIRECTIONS COULD DAMAGE THE EQUIPMENT
11/2014
FUSE TYPE SHOULD BE: 5 x 20 mm., et
0.5 A
F 250 V
Page 9

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido