Problem
Incorrect level readings
Replacement parts and accessories
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Clip, desiccant cartridge
Desiccant beads, bulk
Desiccant cartridge assembly
Filter, hydrophobic
O-ring, dessicant container end cap, 3.0 cm (1.176 inch) ID x
0.18 cm (0.070 inch) OD
Mounting bracket
Screw, for mounting bracket, 10-32 × 0.375, stainless steel
Pipe mounting kit
Accessories
Description
Mounting plate, bubbler line
Sun shield
Tubing, vinyl, 0.317-cm (1/8-inch) ID, 7.6 m (25 ft)
Tubing, vinyl, 0.317-cm (1/8-inch) ID, 30.5 m (100 ft)
Tubing, vinyl, 0.317-cm (1/8-inch) ID, 152 m (500 ft)
Tubing, stainless steel, 0.317-cm (1/8-inch) ID, 0.61 m (2 ft) (use with vinyl tubing)
Table 1 Troubleshooting (continued)
Possible cause
The bubbler calibration is
incorrect for the site
conditions.
There is a blockage in the
bubbler line.
The desiccant is clogged.
The bubbler line is not
installed correctly in the flow
stream.
W A R N I N G
Solution
Calibrate the bubbler.
Manually purge the bubbler line. If the problem
continues, examine the full length of the bubbler line
for a blockage. If the blockage is not easily removed,
install the full length of new bubbler line. Do not
replace only a section of the bubbler line.
Replace the desiccant. Refer to
desiccant
on page 14.
Make sure that the bubbler line has the correct
orientation in the flow stream. Refer to
on page 10.
Quantity
1
0.68 kg (1.5 lb)
1
1
1
1
1
1
Replace the
Figure 6
Item no.
5057
8755500
5027
3390
5252
8309300
8315600
8319000
Item no.
2900
8319200
2929
2921
2922
2891
English 17