N Lock
Use when operating regularly on non-ferrous substrates.
The
icon appears and the probe will only use the eddy
current principle to shorten measurement time and extend
N
battery life. Useful when measuring coatings over plated steel.
Cal Memory
The current cal setting (ex. Cal 3) displays in the upper right
corner of the LCD. A stored cal setting can be restored when
you return to a particular part. Cal 1 can be adjusted but never
deleted, and is always made active with factory settings after a
Reset (pg. 4). A new cal setting cannot be created if a batch is
open with readings.
Units
Converts the display from microns to mils and vice versa.
Reset
Reset (menu Reset) restores factory settings and returns the
Gage to a known condition. The following occurs:
- All batches, stored measurements, batch names and screen
captures are erased.
- Calibration adjustments are returned to factory settings.
- Menu settings are returned to the following:
Memory = OFF
Hi Res = OFF
Statistics = OFF
Hi Lo Alarm = OFF
Scan Mode = OFF
Auto Dim = ON
Perform a more thorough Hard Reset as follows:
1. Power down the Gage and wait 5 seconds.
2. Simultaneously press and hold the ( + ) and
buttons until the Reset symbol
This returns the Gage to a known, "out-of-the-box" condition.
(FN ferrous/non-ferrous combination probes only)
(Advanced models only)
Setup Menu
4
Cal Lock = OFF
Bluetooth & Stream = OFF
WiFi & Access Point = OFF
Display = None
N Lock = OFF
appears.