Follow the proper connection as shown in Fig. 1. The LED (red) indicator
will light. This indicates that a correct current flow is taking place.
Battery replacement:
When the symbol "
•
Disconnect the test leads from the instrument and turn off the power.
•
Unscrew the cover, then slide the cover, take out the batteries and replace with new
batteries, type AA.
•
Replace back cover and secure with 2 screws.
•
If the meter is not to be used for periods of longer than 60 days, remove the batteries
and store them separately.
Maintenance:
Clean the case with a damp cloth and mild detergent : do not use abrasives or solvents.
Service and Replacement Parts:
This unit has no user-serviceable parts.
For replacement parts or to inquire about service information, contact IDEAL INDUSTRIES, INC.
at 1-877-201-9005 or visit our website www.testersandmeters.com.
SPECIFICATIONS
Measurement System:
Earth resistance:
Resolution:
Earth voltage:
Accuracy:
Earth voltage:
Display:
Low battery indication:
Data hold indication:
Over range indication:
Safety standard:
Power source:
Dimensions:
Weight:
Accessories included:
" appears on the display, replace with new batteries as follows:
Operating Frequency - 820Hz, 2mA approx.
Range: 20/200/2000Ω
0~19.99Ω (0.01Ω)
0~199.9Ω (0.1Ω)
0~1999Ω (1Ω)
0~200V AC, 40-500Hz
Earth resistance: ±(2%reading± 2digits) or ±0.1Ω,
whichever is greater.
±(1% reading±2digits)
LCD 3½ digit (2000 counts)
"
"Symbol appears on the display.
"
HOLD
" Symbol appears on the display.
" 1 " (MSD).
IEC/EN 61010-1, Installation CATIII - 200V.
(8) 1.5V (AA) batteries required.
8.3"Hx8.3"Wx4.0"D (210mmLx210mmWx100mmD)
Approx. 3 Lbs(1361g) (batteries included).
Hard Case, Test leads (red-33m, yellow-33m, green-5m), 2 Auxiliary
earth spikes, simplified measurement probe, Instruction manual.
3