Comparing the Control Solution Test Results

The test result of each control solution should be within the
range printed on the label of test strip vial. Repeat the control
solution test if the test result falls outside of this range.
Out of range results may occur due to the following factors.
Situations
• When the control solution bottle was not
shaken well
• When the meter, test strip or the control
solution were exposed to high or low
temperatures
• When the first one or two drops of the
control solution were not discarded or the
tip of the vial was not wiped clean
• When the meter is not functioning properly
• When the control solution is past the
expiration date printed on the bottle
• When the control solution is past its
discard date (the date the bottle was
opened plus three (3) months)
• When the control solution is contaminated
Actions
Repeat the
control solution
test by referring
to the 'NOTE' on
page 19.
Discard the used
control solution
and repeat the
test using a new
bottle of control
solution.
23