Calibration Verification - Hach FT660 sc Manual Básico Del Usuario

Ocultar thumbs Ver también para FT660 sc:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

7. Put the bubble trap cover and head assembly on the body of the
instrument.
Note: If the head assembly is not correctly installed, the laser will not turn on
and the reading on the display will be near 0.000 mNTU or a negative value.
The red status light on the instrument will also flash one time per second.
8. When the turbidity reading is stable (and after greater than
5 minutes), confirm to accept the reading.
9. When the display shows "GOOD CAL! GAIN: X.XX ENTER TO
CONT.", confirm to save the turbidity value.
Note: If the display shows "BAD CAL!", confirm to do the calibration again or
exit.
10. When the display shows "VERIFY CAL?", confirm to do a calibration
verification or exit without verification.
11. Do a calibration verification or enter the operator initials to complete
the calibration.
12. When the display shows "RETURN SENSOR TO MEASURE
MODE", confirm.
Figure 7 Calibrate the instrument
1 Head assembly
®
2 800 mNTU StablCal
Certified
Standard
3 Bubble trap

Calibration verification

Do a calibration verification immediately after every calibration.
Subsequent verifications are referenced to the initial verification.
Do a calibration verification monthly between calibrations.
A calibration verification is a check that is done to make sure that the
instrument is working correctly and calibrated.
4 Funnel
5 Body of the instrument
6 Bubble trap cover
English 11

Publicidad

Tabla de contenido
loading

Tabla de contenido