Operation

Cure Memory function

The last settings used, together with the combination of
curing program and curing time, are automatically saved.

Start

The light is switched on by means of the start button. Once
the selected curing time has elapsed, the curing program is
automatically terminated. If desired, the light can be
switched off before the set curing time has elapsed by
pressing the start button again. The fan is activated simulta-
neously to the light. Once the curing time has elapsed, the
fan continues to run for a certain time to cool the appara-
tus. The battery must not be removed as long as the fan is
still running.

Acoustic signals

Acoustic signals can be heard for the following functions:
• Start (Stop)
• Every 10 seconds
• Program change
• Curing time change
• Connecting the battery to the charging base
• Inserting battery
• Error message
If desired, the volume of the acoustic signals can be
adjusted.
For this purpose, press the blue volume button (see Product
overview) while the light is off and reduce the volume or
turn the acoustic signals off with the left time selection but-
ton. If you want to turn the acoustic signals back on or
increase the volume, press the blue volume button and then
the right time selection button.
16

Light intensity

The light intensity is maintained at a consistent level during
operation. If the supplied 10>8 mm light probe is used, the
light intensity has been calibrated to 2,000 - 2,200 mW/cm

Measuring the light intensity

The light intensity of the bluephase 20i and the supplied
10>8 mm light probe can be checked, for instance, by
means of the bluephase meter.
If the measured value does not correspond with the expect-
ed light intensity, proceed as follows:
• Check the selected curing program
• Clean light sensor if it is contaminated
• Remove the light probe and clean the light emission
window of the handpiece with a cotton swab dipped in
alcohol
• Clean light probe if it is contaminated
(see Maintenance and cleaning)
• Replace a damaged light probe with a new one
If these measures do not improve the results, please contact
your dealer or your local Service Center.