ZEISS
TECHNICAL DATA
Axio Observer
7
TECHNICAL DATA
Dimensions (width x depth x height)
Stand Axio Observer 3, 3 materials;
........................................................................................................
Stand Axio Observer 5, 5 materials and
....................................................................................................
Stand Axio Observer 7, 7 materials ....................................... approx. 295 mm x 805 mm x max. 707 mm
Weight
Axio Observer 3, 3 materials .............................................................................................. approx. 27 kg
Axio Observer 5, 5 materials .............................................................................................. approx. 30 kg
Axio Observer 7, 7 materials .............................................................................................. approx. 36 kg
Ambient conditions
Storage (in packaging)
Permissible ambient temperature .................................................................................. +10 °C to +40 °C
Permissible air humidity (without condensation) ...................................................... max. 75 % at +40 °C
Transport (in packaging):
Permissible ambient temperature ................................................................................... -40 °C to +70 °C
Permissible air humidity (without condensation) ...................................................... max. 75 % at +35 °C
Operation
Permissible ambient temperature ......................................................... +10 °C to +35 °C, optimally 22 °C
Permissible air humidity ............................................................................................. max. 65 % at 30 °C
Air pressure ............................................................................................................ 800 hPa to 1060 hPa
Degree of pollution ............................................................................................................................... 2
Altitude of operating site ..................................................................................................... max. 2000 m
Operating data for Axio Observer 3, 3 materials and 5, 5 materials with integrated power
supply unit, and Axio Observer 7, 7 materials with external power supply unit VP232-2
Operating area ............................................................................................................... Enclosed rooms
Protection class ..................................................................................................................................... I
Ingress protection rating ................................................................................................................. IP 20
Electrical safety ........................................................................................ DIN EN 61010-1 (IEC 61010-1)
.............................................................................................
and conforming to CSA and UL regulations
Overvoltage category ............................................................................................................................ II
Suppression of interference ............................................................ in accordance with EN 55011 Class B
Noise immunity ................................................................................ in accordance with DIN EN 61326-1
Line voltage ...................................................................... 100 V to 127 V and 200 V to 240 VAC ±10 %
(Axio Observer 3, 3 materials / 5, 5 materials)
8