Heathrow Scientific Magnetic Manual De Instrucciones página 8

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Alternating Timed Mixing
9.
Press knob to start unit.
10. Turn knob to right to view "Reverse" mode for multi-direction
spinning. Press knob to select this mode.
11. Turn knob to desired speed setting. Press knob to select rpm.
12. Turn knob to select reverse time (00:30 to 60:00) to set time cycle for
spinning direction. Press knob to select time cycle.
13. Turn knob to desired number of cycles (01 to 99).
14. Press knob to start spin cycling.
Note: When adjusting speed while unit is running, the display will
show the set speed. Once set, the display will show the ramping
speed until the stir speed is reached.
Note: If the speed adjustment is done rapidly, the stir bar may
uncouple and jump around erratically. Press the knob to stop, move
the container to recouple stir bar, and ramp up the speed slowly.
Note: Settings are retained for next use when power is turned off.
15. Once cycles are complete, stirring will stop. Then you can remove
the mixing container.
Note: To cancel set program, press and hold knob for 3 seconds.
Warning!
Do not drop stir bars into empty glass containers as the container may
break.
Warning!
Do not plug other power adaptors into stirrer unit, this will void the
warranty.
Warning!
Do not use if cord or power supply is damaged.
Warning!
Do not quickly turn speed knob as the liquid may spill out of container.
Maintenance
The Magnetic Induction Stirrer is a maintenance free product. Clean unit
only when it not plugged into an electrical outlet. When necessary, the
complete outside of the unit can be wiped down using a damp cloth
and a mild, pH neutral detergent. Do not immerse the unit or the power
adapter.
8

Publicidad

Tabla de contenido
loading

Tabla de contenido