Safety Information
1-2
Cryogenic Transfer Vessels
Warning The venting of nitrogen vapors will create a dilution of the air's
oxygen concentration necessary to support life. Exposure to this diluted
atmosphere can cause asphyxiation or even death. To avoid personal
injury or death, DO NOT store or use liquid container in areas that have
poor ventilation. Use in a well ventilated area. Place liquid containers
outdoors or in a well ventilated area. Failure to comply with this warning
may cause serious personal injury or even death.
Warning Extreme temperature and pressure are associated with the
Thermo 8127. To avoid serious personal injury, observe caution when
removing parts or fittings. Wear eye protection and insulated, loose
fittings gloves when removing parts or fittings. Failure to comply with
this warning may result in serious personal injury.
Caution DO NOT use regulators, valves, gauges, hoses, etc. that have been
used in compressed air service. Failure to observe this warning could cause
serious damage to your container and possible personal injury.