Using the hand control
Session Time
This function allows you to see how long the unit has been operating since it was last reset.
Beginning
To stop the unit
Press simultaneously on the 2 keys
To reset the counter, press the key
Or
After 5 seconds the time is no longer displayed, but it
remains in the memory.
Quick Start
Set up the patient and proceed as below:
Beginning
Switch the unit ON
To choose the movement
To start the session with the original parameters for the
movement (default setting)
Possible values for each parameter
• Lower limit
• Upper limit
• Speed
• Force
• Extension pause
• Flexion pause
• Timer
Kinetec Centura B&W™
Keys to
press
Keys to
press
Flexion/Extension - arm straight
Possible values
0° to 105°
5° to 110°
1 to 5 (from 45° to 130° per
minute)
1 to 6
0 to 900 seconds (15 minutes)
0 to 900 seconds (15 minutes)
No timer (00H00) to 24H00
Display
ARM FLEXED
50
STOP
0
TOTAL TIME 0125H
RESET: low.limit
TOTAL TIME 125H
Reset to zero
ARM FLEXED
50
STOP
0
Display
KINETEC
Centura B&W V01.0
Movement verif.
Please wait
Arm straight
30
STOP
0
ARM FLEXED
50
STOP
0
Arm straight
30
STOP
0
Arm straight
30
STOP
0
Default setting
30°
90°
1 to 5 (from 45° to 130° per
2
6
0
0 to 900 seconds (15 minutes)
0
0 to 900 seconds (15 minutes)
No timer
No timer (00H00) to 24H00
10/24
Remarks
Check that the hand control is not locked (See
90
page 9).
The display shows how long the unit has been
operating since it was last reset.
This message disappears after a few seconds;
the counter is now reset.
90
Remarks
Check that the hand control is not locked (See
page 9).
90
90
The display shows the movement selected, with
the default settings.
90
The angle display changes with the current
90
movement.
Flexion/Extension - arm flexed
Possible values
Default setting
0° to 105°
5° to 110°
minute)
1 to 6
30°
90°
2
6
0
0
No timer
EN