This symbol on the instrument indicates that the user should refer to the
manual for operating instructions.
The vial port contains sharp needles designed to pierce the septa of sample
vials. Do not put fingers or inappropriate materials into the vial port.
To protect against inhalation of toxic fumes, make sure the Analyzer is installed
in a well-ventilated area. To treat the gas waste stream, a halogen trap is
provided and will shut down the instrument when the trap has expired.
Depending on the composition of the sample, halogens such as chlorine or
bromine, or toxic fumes from fluoride, cyanide, or other toxic salts, can be
emitted into the Analyzer's liquid waste stream. To protect against inhalation of
hazardous fumes, add sodium hydroxide (lye) to the waste container to
neutralize acid in the waste. Avoid stratification of the waste within the
container by occasionally agitating the container.
This symbol indicates that to comply with European Union Directive 2002/96/
EC for waste electrical and electronic equipment (WEEE), the Analyzer should
be disposed of separately from standard waste.
For continued protection against fire hazard, replace the fuse with a fuse of the
same type and rating.
SUEZ © 2018
Warning
Warning
Warning
Warning
Warning