Guidance and manufacture's declaration – electromagnetic immunity
The ri-former
Diagnostic Station is intended for use in the electro-
®
magnetic environment specified below. The customer or the user of the
Diagnostic station ri-former should assure that it is used in such an
environment.
IEC 60601
Compliance
Immunity test
test level
level
Conducted RF
IEC 61000-4-6
Pass
Pass
Radiated RF
3 V/m
3 V/m
IEC 61000-4-3
80 MHz to
2,7 GHz
Proximity fields
Pass
Pass
from RF wireless
communi-cations
equipment
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a.)
Field strengths from fixed transmitters, such as base stations for radio (cellular/cord-
less) telephones and landmobile radios, amateur radio, AM and FM radio broadcast and
TV broadcast cannot be predicted theoreticallywith accuracy. To assess the electroma-
gnetic environment due to fixed RF transmitters, an electromagnetic sitesurvey should
be considered. If the measured field strength in the location in which the Diagnostic
station ri-former is used exceeds the applicable RF compliance level above, the Dia-
gnostic station ri-former should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting
or relocating the ri-former
®
Diagnostic Station.
b.)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between
portable and mobile RF communications equipment and the
ri-former
The
ri-former
gnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the
prevent electromagnetic interference by maintaining a minimum distance
Electromagnetic
between portable and mobile RF communications equipment (trans-
environment - guidance
mitters) and the
according to the maximum output power of the communications equip-
Portable and mobile RF
ment.
communications equipment
should not be used closer to
Rated maximum
any part of the ri-former
®
Diagnostic Station, including
output power of
the cables, than the recom-
transmitter
mended distance, which is
calculated using the equation
(W)
applicable to the transmitter
frequency.
Recommended separation
distance
d= 1.2√P 150 KHz to 80 MHz
0,01
d= 1.2√P 80 MHz to 800 MHz
d= 2.3√P 800 MHz to 2,7 GHz
0,1
1
Where P is the maximum
output power rating of the
10
transmitter in watts (W)
according to the transmitter
100
manufacturer and d is the
For transmitters rated at a maximum output power not listed above, the recommended
recommended separation
separation distance d inmetres (m) can be estimated using the equation applicable to the
distance in metres (m).
frequency of the transmitter, where P is themaximum output power rating of the trans-
Field strengths from fixed RF
mitter in watts (W) according to the transmitter manufacturer.
transmitters, as determined
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
by an electromagnetic site
survey,a should be less than
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
the compliance level in each
frequency range.b
Interference may occur in the
vicinity of equipment marked
with the following symbol:

11. Accessories

Where P is the maximum out-
put power of the transmitter
in watts (W) according to the
Art. No.: 3652-600
transmitter manufacturer and
Art. No.: 3652-500
the recommended distance is
given in meters (m).
Art. No.: 3652-400
Field strengths from fixed RF
transmitters determined by
Art. No.: 10384
an electromagnetic site
survey a, should be less than
the compliance level in each
frequency range b.
Art. No.: 3655-103
Interference may occur in the
Art. No.: 3655-106
vicinity of devices marked with
the following symbol:
Art. No.: 3655-109
Art. No.: 3655-123
Art. No.: 3655-130
Art. No.: 3656
Art. No.: 3656-301
Art. No.: 3654
Art. No.: 14065-531
Art. No.: 14065-534
Art. No.: 10563
Art. No.: 10563-301
Art. No.: 10580
Art. No.: 10580-301
Art. No.: 10565
Art. No.: 10565-301
Diagnostic Station.
®
Diagnostic Station is intended for use in an electroma-
®
Diagnostic Station can help
ri-former
®
Diagnostic Station as recommended below,
ri-former
®
Separation distance according to frequency of
transmitter
(m)
150 kHz to 80
80 MHz to 800
MHz
MHz
d =12√P
d =1,2√P
0,12
0,12
0,38
0,38
1,2
1,2
3,8
3,8
12
12
applies.
affected by absorption andreflection from structures, objects and people.
ri-former
with foot and big ben® 3.5 V / 100-240 V
®
ri-former
®
with foot 3,5 V / 100-240 V
ri-former
anesthesia model without universal
®
clamp 3.5 V / 100-240 V
Universal clamp
Extension module, big-ben
big ben® Wickelman. adult
Hakenman. adult
Klettenman. adult
Klettenman. strong arms
Klettenman. children
Extension module, infrared thermometer
ri-thermo®N
Extension module, infrared thermometer ri-ther-
mo®N without anti-theft device
Extension module, infrared thermometer ri-ther-
mo®N with anti-theft device
ri-spec Ear specula Dispenser Packed in a polybag
25 pcs.
1 box 40 poly bags 1000 pcs. 2.5 mm hopper Packed in
a polybag 25 pcs.
1 box 40 poly bags 1000 pcs. 4 mm funnels Packed in
a polybag 25 pcs.
ri-former
Instrument heads
®
without anti-theft device with anti-theft device
ri-scope® L F.O. Otoscope
L1 3,5 V XL
L1 3,5 V XL with anti-theft device
L2 3,5 V XL
L2 3,5 V XL with anti-theft device
L2 3,5 V LED
L2 3,5 V LED with anti-theft device
800 MHz to
2,7 GHz
d =2,3√P
0,23
0,73
2,3
7,3
23
23