observed to verify normal operation in the configuration in which it will be used.
Guidance and manufacture's declaration – electromagnetic emission
The device is intended for use in the electromagnetic environment specified below. The
customer of the user of the device should ensure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emission
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/ flicker
emissions
IEC 61000-3-3
Guidance and manufacture's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The
customer or the user of the device should ensure that it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
Compliance
Group 1
Class B
Class A
Complies
IEC 60601 test
level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for
input/output lines
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
Electromagnetic environment – guidance
The device uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
The device is suitable for use in all
establishments, including domestic
establishments and those directly connected to
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
Compliance
level
±6 kV contact
±8 kV air
±2kV for power
supply lines
±1 kV differential
mode
<5% U
T
(>95% dip in U
)
T
for 0.5 cycle
40% U
T
(60% dip in U
)
T
for 5 cycles
7
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the device requires
continued operation during
power mains interruptions, it
is recommended that the