SECTION 7—CLEANING, CARE, AND MAINTENANCE
Air Intake Filter
DO NOT operate the XPO
installed.
NOTE: For this procedure, refer to FIGURE 7.1.
NOTE: Remove the air intake filter and clean it at least once each
week.
1. Lift filter cover slightly and pull down to remove tabs from
the grooves.
2. Lift out the filter.
3. Use a vacuum cleaner or wash with a mild liquid dish
detergent (such as Dawn™) and water. Rinse thoroughly.
4. Thoroughly dry the filter and inspect for fraying,
crumbling, tears and holes. Replace filter if any is found.
5. Reinstall the filter cover by placing the tabs in the slots and
engaging the magnet.
NOTE: Use only Invacare part number 1148281 as the air intake
filter for your XPO
Cover
Invacare® XPO
™
2
CAUTION
without the air intake filter
2
.
2
Magnet
FIGURE 7.1 Air Intake Filter
Grooves
Cover
38
Tabs
Magnet
Filter
Part No 1160887