SECTION 8—TROUBLESHOOTING GUIDE
ALARM/WARNING:
SYMPTOM
BREATH RATE OVER CAPACITY ALARM:
Intermittent
audible beep and
illuminated red
alarm indicator.
Audible beeps and
red indicator blinks
become more
rapid as breathing
rate increases.
SYSTEM TOO HOT/COLD TO START ALARM:
Continuous
audible beep and
red alarm indicator
is illuminated. Fan
is On.
AND
Flow indicators 1
& 2 are
illuminated.
SYSTEM RUNNING TOO HOT/COLD ALARM:
Continuous
audible beep and
red alarm indicator
is illuminated. Fan
is On.
AND
Flow indicators 1
& 3 are
illuminated.
Invacare® XPO
™
2
PROBABLE CAUSE SOLUTION
Your breathing rate has
exceeded the capacity of the
XPO
.
2
Unit is too hot, or too cold,
to allow it to turn On.
Unit has become too hot, or
too cold, during operation.
42
1. Immediately reduce
your activity level to get
your breath rate to slow
down, and / or
2. If the alarm continues,
change to another
source of oxygen.
1. Allow unit to cool
down to less than 95° F
(35° C), or warm up to
50° F (10° C), before
turning the unit Off,
then On again.
2. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider
1. Allow unit to cool
down to less than 95° F
(35° C), or warm up to
50° F (10° C), before
turning the unit On again
2. If the alarm continues,
change to another
source of oxygen and
contact your equipment
provider
Part No 1160887