TM
Frontier
Centrifuge FC5513
4.
SETTING
4.1
Basic adjustments
4.1.1
Change the type of rotor
Before the first operation and after each rotor change, you have to set the rotor type. You find each rotor type in the
printed order number on the rotor.
Example:
Angle rotor order number: 30130871
Rotor Type on the display = 71
Turn on the centrifuge and open the lid. Now simultaneously press the keys "Quick Spin"(8) and "Stop/Lid"(9). In
the display "RPM/RCF" the old rotor type no "71" appears. With the potentiometer you can now set the rotor type.
To store the new setting please press the "Start" (10) key. Inside the display, "Store" appears as confirmation.
Now all important rotor parameters for the centrifuge are stored.
ATTENTION:
The set rotor type must always be the same as the actual rotor type used; otherwise the equipment might
be damaged.
The rotor type can be checked during the run by pressing the key "Quick Spin"(8).
4.1.2
Access to mode "Operating Data"
When using the centrifuge, the following parameters can be set:
- Acoustic signal turn on/off
- Keyboard sound turn on/off
M11
Figure. 26
A-1
Figure. 27
A-4
2
10
9
8
EN-13