4.2.
Electrical specifications
Accuracy :
Resistance
Frequency test : 100 Hz / 120 Hz
Range
10 MΩ
2000 kΩ
200 kΩ
20 kΩ
2000 Ω
200 Ω
20 Ω
Frequency test : 1 kHz / 10 kHz
Range
10 MΩ
2000 kΩ
200 kΩ
20 kΩ
2000 Ω
200 Ω
20 Ω
Notes
IX 3131B
@ 23° C ± 5° C ; < 75 % H. R.
Max.
Display
9.999 MΩ
1999.9 kΩ
199.99 kΩ
19.999 kΩ
1999.9 Ω
199.99 Ω
19.999 Ω
Max.
Display
9.999 MΩ
1999.9 kΩ
199.99 kΩ
19.999 kΩ
1999.9 Ω
199.99 Ω
19.999 Ω
1 - The components under test and measurement cords
must be properly shielded or linked to the earth terminal
if necessary.
2 - In this range, the accuracy is specified in parallel mode.
Accuracy
100 Hz / 120 Hz
0.6 % + 5
0.3 % + 3
0.3 % + 2
0.3 % + 2
0.3 % + 2
0.5 % + 3
0.6 % + 40
Accuracy
1 kHz
10 kHz
0.6 % + 5
2.5 % + 10 *N2
0.3 % + 5
0.8 % + 10 *N2
0.3 % + 2
0.6 % + 5
0.3 % + 2
0.6 % + 5
0.3 % + 2
0.6 % + 5
0.5 % + 3
1.2 % + 25
0.6 % + 40
1.2 % + 200
Notes
after open cal.
after open cal.
-
-
-
after short cal.
after short cal.
Notes
after open cal.
after open cal.
-
-
-
after short cal.
after short cal.
II - 41