Error Messages
MESSAGE
Err 1
Inflation or pressure error.
Err 2
Err 3
Cuff pressure is too high.
Err 4
No pulse rate detected or
Err 5
the pulse rate is too low.
Systolic pressure is
Err 6
outside the system s
measurement range
Forced to stop the blood
Err 8
pressure measurement
after re-pump for 2 times.
Err 15
Data corrupted.
Appears when the
batteries cannot provide
enough power for a test.
E-U
Strip has been used.
Environmental
E-t HIGH
temperature is outside
E-t LOW
the system s operational
range.
E-d
Problem in operation.
WHAT IT MEANS
WHAT TO DO
Refit cuff tightly and correctly. Relax
and repeat the measurement. If the
error still remains, please contact local
customer service for help.
System measurement range is 60
mmHg – 250 mmHg. Review the
instructions and repeat the test. If the
error still remains, please contact your
healthcare professional for help.
Turn off the device and start over.
Follow the instructions and try again.
Follow the instructions and try again.
If problem persists, contact the local
customer service for assistance.
Replace or recharge the batteries
immediately.
Repeat the test with a new strip.
System operational range is 50 ° F to
104 ° F (10 ° C to 40 ° C). Repeat the
test after the meter and test strip have
reached the above temperature.
Review the instructions and repeat the
test with a new strip.
49