Caring for Your Meter
Wash and dry your hands thoroughly before use or handling of the
meter and strips.
What is Cleaning and Disinfection?
Cleaning and disinfection are different. Cleaning is the process of
removing dirt (e.g. food debris, grease, dust), disinfection is the process
of killing germs (e.g. bacteria and viruses).
When to clean and disinfect the meter?
Clean the meter when you see any dirt on it. Disinfect the meter
between each patient to prevent infection.
How to clean and disinfect the meter?
The meter must be cleaned prior to the disinfection. Use one
disinfecting wipe to clean exposed surfaces of the meter thoroughly and
remove any visible dirt or blood or any other body fluid with the wipe.
Use a second wipe to disinfect the meter. Do NOT use organic solvents
to clean the meter.
We recommend for meter cleaning and disinfection you should use the
disinfecting wipe/towelette from below. The following product has been
shown to be safe for use with the FORA D40 Pro blood glucose
motoring system.
► Micro-Kill+™ (Micro-Kill Plus™) by Medline
(EPA Reg. No. 59894-10-37549)
To obtain disinfecting wipes and other information, please contact
Medline at 1-800-MEDLINE (1-800-633-5463) or visit
www.medline.com.
32