Each cleaning and disinfection cycle includes a pre-cleaning step with
one wipe and a disinfection step with a second wipe. Improper system
cleaning and disinfection may result in meter malfunction. If you have a
question, please contact customer service at 1-888-307-8188 for
assistance.
This device has been validated to withstand 260 cleaning and
disinfection cycles using the recommended disinfecting wipe/towelette.
The tested number of cycles is estimated by 1 cleaning and disinfection
cycle per meter per week for 5 years, the expect life of device.
Stop using the meter if you see any signs of deterioration such as the
LCD display is cracked or cloudy, the buttons no longer function, or the
outer casing is cracked. Please contact customer service at
1-888-307-8188 for a replacement meter if any of the signs of
deterioration are noticed.
The meter should be replaced after the validated number of disinfection
cycles or the warranty period, whichever comes first.
Stop using the meter if you see any signs of deterioration. For example:
meter can not be turned on,
LCD display cracks or becomes cloudy,
buttons no longer function,
meter outer casing cracks,
data can not be transmitted to PC,
color or paint/printing on housing is abnormal, or
scratches or abrasions on meter are higher than acceptable.
Please contact the customer service for a replacement meter if any of
the signs of deterioration are noticed.
34