measure. Then the Abs. value would appear in the corresponding
table.
WL: 500.0nm
Calibration Table
No.
1
2
3
C=0.226+1.031*A+^1
Method
Load the Stored Curves
In the "Calibration Table" interface, press
select interface. Use
EN TER
to load. Users can press
WL: 500.0nm
Calibration Table
No.
Conc.(mg/ml)
1
0.000
2
1.000
3
2.000
C=0.226+1.031*A+^1
Method
Step 4.
Return the sample measurement interface
In the "Calibration Table" interface, press
Quantitative Test interface.
WL: 500.0nm
Calibration Table
No.
Conc.(mg/ml)
1
2
3
C=0.226+1.031*A+^1
Method
Step 5.
Measure Samples
Place the sample to be tested in the light path, press
measure. Then the test result will display in the data sheet.
Repeat this step to finish measuring all the samples.
10:00:00
D2
W
Conc.(mg/ml)
Abs
WL.(nm)
0.000
1.000
500.0
2.000
r=0.962
Params
Standard
Show Curve
,
to select the curve you need and press
10:00:00
D2
W
Abs
WL.(nm)
0.102
0.200
500.0
0.297
r=0.999
Params
Standard
Show Curve
10:00:00
D2
W
Abs
0.000
0.102
WL.(nm)
1.000
0.200
500.0
2.000
0.297
r=0.999
Params
Standard
Show Curve
WL: 500.0nm
Calibration Table
No.
Conc.(mg/ml)
1
0.000
2
1.000
3
2.000
C=0.226+1.031*A+^1
r=0.999
Method
Params
OPEN
to go into files
to view the curve.
WL: 500.0nm
0.04
Abs
Press
ESC
to return...
Params
ES C
WL: 500.0nm
Quantitative Test
No.
Abs
Conc.(mg/ml)
C=0.226+1.031*A+^1
r=0.999
Unit
Fit Curve
10:00:00
D2
W
Abs
WL.(nm)
0.102
0.200
500.0
0.297
Standard
Show Curve
10:00:00
D2
W
WL.(nm)
500.0
0.46
to return the
10:00:00
D2
W
WL.(nm)
500.0
Search
Scroll
S TART
S TOP
to