Metrohm 719 S Titrino Instrucciones Para El Uso página 76

Tabla de contenido

Publicidad

3.2. Remote control commands
Object
& Root
:
Assembly
. Bur
│ ├
.Rates
│ │ ├
.Forward
│ │ │ ├
.Select
│ │ │ ├
.Digital
│ │ ├
.Reverse
│ │ │ ├
.Select
│ │ │ ├
.Digital
│ ├
.Fill
│ ├
.ModeDis
│ │ ├
.Select
│ │ ├
.V
│ │ ├
.Time
│ │ ├
.VStop
│ │ ├
.AutoFill
.Meas
│ ├
.Status
│ ├
.MeasInput
│ ├
.Ipol
│ ├
.Upol
.Outputs
│ ├
.AutoEOD
│ ├
.SetLines
│ │ ├
.L0
│ │ ├
up to L 3
│ ├
.ResetLines
72
&Assembly
Description
Assembly control
Buret
Rates
Forward rate
Type of rate control
Digital rate
as for forward rate
Type of rate control
Digital rate
Fill
Dispensing
Type of dispensing control
Volume to be dispensed
Time to dispense
Limit volume
Filling after each increment
Measuring
Measuring ON/OFF
Selection of measuring input
Polarization current
Polarization voltage
I/O outputs
Automatic output of EOD
Set I/O lines
Signal on L0
Reset I/O lines
Input range
digital, analog
0...150, max.
digital, analog
0...150, max.
$G,$H,$C
$G,$S,$H,$C
volume, time
0.0001...0.1...9999
0.25...1...86 400
0.0001...9999, OFF
ON, OFF
ON, OFF
1, 2, diff., Ipol, Upol, Temp
0...1...±127
0...400...±1270
ON, OFF
$G
active,inactive,pulse,OFF
$G
Reference
3.2.2.72.
ditto
ditto
ditto
3.2.2.73.
3.2.2.74.
ditto
ditto
ditto
ditto
ditto
3.2.2.75.
ditto
ditto
ditto
3.2.2.76.
ditto
ditto
ditto
719 S Titrino

Publicidad

Tabla de contenido
loading

Tabla de contenido