Guidance and manufacturer's declaration – electromagnetic immunity
The device is inteded for use in the electromagnetic environment specified below. The customer or the
user of the device should assure that is used in suche environment
Immunity test
Test level EN
Conducted RF
3 Vrms 150kHz
EN 61000-4-6
to 80MHz
RF Radiata
3 Vrms 80MHz
EN 61000-4-3
to 2,5GHz
Field strangths from fixed RF transmitters, are determined by an electromagnetic site survey, should
be less than the complicance level in each frequency rage.
Interference may occur in the vicinity of equipment marked with the following symbol:
Recommended separation distances between portable and mobile communications equipment
The device is intended for the use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent electromagnetic
interferences by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the device as recommended below, according to the maximum output
power of the communication equipment.
Rated
maximum
power
of
the
transmitter (W)
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in Watts (W) according to
the transmitter manufacturer.
Note:
(1) At 80 MHz and 800 MHz the separation distance for the higher frequency range applies.
(2) These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
I.A.C.E.R. Srl
Compliance level
60601-1-2
3 Vrms 150kHz to
80MHz
3 Vrms 80MHz to
2,5GHz
and the device
Separation distance according to the frequency of the transmitter (m)
150kHz to 80MHz
d = 1,2 P
0,12
0,38
1,2
3,8
12
44de 48
Electromagnetic environment –
guidance
Portable
and
communications equipment should be
used no closer to any part of the device,
including
cables,
recommended
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 P 150kHz to 80MHz
d = 1,2 P 80 MHz to 800 MHz
d = 2,3 P 800 MHz to 2,5 GHz
where (P) is the maximum output power
rating of the transmitter in Watts (W)
according
to
manufacturer
recommended separation distance in
metres (m).
80MHz to 800MHz
d = 1,2 P
0,12
0,38
1,2
3,8
12
MNPG158-06 27/06/2016
mobile
RF
than
the
separation
distance
the
transmitter
and
(d)
is
the
800MHz to 2GHz
d = 2,3 P
0,23
0,73
2,3
7,3
23

Publicidad