e. Do not touch bulb with fingers, use a clean cloth and insert pins of bulb into socket.
f.
If bulb has been touched, clean it as this could affect light transmission.
g. Close the cap, and screw down firmly.
•
For the SFC-100FLED
a. Lift up the stage using the focusing knobs.
b. Unscrew the screws indicated. (Fig. 2)
c. Take out the top part of the illuminator carefully
without damage the bulb, the condenser or the
illuminator.
d. With carefully take out of the bulb and pull
outwards to disconnect it from the socket.
e. Do not touch the new bulb with bare hands.
Use a clean cloth to insert the bulb pins into the
socket.
f.
If the bulb is touched with bare hands, it must
be cleaned, as the transmission of light could
be affected.
g. Put on again the top part of the illuminator and tight the screws.
2. Changing the fuse (Fig. 3).
a. With a flat screwdriver, press lightly on the slot on
the fuse holder cap and turn ¼ way in the direction
of the arrow indicated.
b. Release pressure and remove cap.
c. Remove the fuse from the cap by pulling it out, and
insert a new one, ensuring that it is 0.5 Amps.
d. Insert cap with fuse.
e. Repeat (a.) turning 1/4 in the opposite direction to
the arrow indicated. The cap should be firmly
closed.
-7-
Fig. 2
Fuse
Fig. 3