27 7.1 Change Of Location; 27 7.2 Disposal And Scrapping; Technical Data - Bosch FSA 050 Manual Original

Ocultar thumbs Ver también para FSA 050:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 89
7.
Decommissioning
7.1
Change of location
If the FSA 050 is passed on, all the documentation
included in the scope of delivery must be handed over
together with the unit.
The FSA 050 is only ever to be transported in the
original or equivalent packaging.
Unplug the electrical connection.
Heed the notes on initial commissioning.
7.2
Disposal and scrapping
FSA 050, accessories and packaging should
be sorted for environmental-friendly recycling.
Do not dispose FSA 050 into household
waste.
Only for EC countries:
The FSA 050 is subject to the European
directive 2012/19/EC (WEEE).
Dispose of used electrical and electronic
devices, including cables, accessories and
batteries, separately from household waste.
Make use of the local return and collection
systems for disposal.
Proper disposal of FSA 050 prevents
environmental pollution and possible health
hazards.
Robert Bosch GmbH
Decommissioning | FSA 050 | 35
8.

Technical data

i
All the values given are referenced to a temperature
of 20 °C.
Insulation analysis
i
The specifications only apply in conjunction with
the remote sensor and the measurement leads
contained in the scope of delivery.
Range
Accuracy
1000 V
200 GΩ ±3 % ±2 positions ±0.2 % per GΩ
500 V
100 GΩ ±3 % ±2 positions ±0.4 % per GΩ
250 V
50 GΩ ±3 % ±2 positions ±0.8 % per GΩ
100 V
20 GΩ ±3 % ±2 positions ±2.0 % per GΩ
50 V
10 GΩ ±3 % ±2 positions ±4.0 % per GΩ
Function
Specification
Analog
1 GΩ for full scale deflection
display range
Short-circuit
2 mA 0 % – 50 %
current
Terminal
0 % – 20 % ±1 V (Ii <1 mA)
voltage
Test current level
1 mA for minimum insulation continuity
under load
value according to specification as per
BS7671, HD384 and IEC364, EN 61152-2,
2 mA maximum
EN 61557
0.10 MΩ to 1.00 GΩ
operating range
Leakage current
10 μA to 2000 μA
measurement
range
Leakage current
10 % ±3 positions
Voltage display
3 % ±3 positions ±0.5 % of rated voltage
Polarisation
Ratio of 10 minutes to 1 minute
index (PI)
Dielectric
Ratio of 60 s to 30 s
absorption ratio
(DAR)
Continuity measurement
Function
Specification
EN 61557
0.01 Ω to 99.9 Ω
operating range
(0 to 100 Ω on analog scale)
Accuracy
±2 % ±2 positions (0 to 100 Ω)
Voltage in
5 V ± 1 V
open circuit:
Test current
205 mA (±5 mA) (0.01 Ω to 9.99 Ω)
20 mA (±1 mA) (10.0 Ω to 99.9 Ω)
Zero adjustment
Usually 0.10 Ω
at remote
sensor tips
Zero adjustment
Up to 9.99 Ω
of line resistance
Buzzer
Variable limit: 1 Ω, 2 Ω, 5 Ω, 10 Ω, 20 Ω
en
|
1 689 989 102
2019-05-24

Publicidad

Tabla de contenido
loading

Tabla de contenido