Possible Problems and solutions
Problem
SpO
or PR cannot
2
be shown normally.
GB
SpO
or PR is
2
shown unstably
The oximeter can-
not be switched on
Possible Reason
1. Finger is not
inserted cor-
rectly.
2. Patient's
oxyhaemoglobin
level is too low
to be measured.
1. Finger might not
be inserted far
enough.
2. Finger is
trembling or the
patient's body is
moving.
1. Battery power
may be inad-
equate or not be
there at all.
2. Batteries may
have been
inserted incor-
rectly.
3. The oximeter
might be dam-
aged.
28
Solution
1. Try inserting the
fi nger again.
2. Try several more
times if you are
sure that there
is no problem
with the device.
Please go to
a hospital in
good time to
obtain a precise
diagnosis.
1. Try inserting the
fi nger again.
2. Try not to move.
1. Please replace
batteries.
2. Please reinsert
the batteries.
3. Have the device
tested by the
manufacturer