Guidance and manufacturer´s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should
only be used in such environments:
Immunity
test
Conducted
RF IEC
61000-4-6
Radiated RF
IEC 61000-
4-3
GB
NOTE 1:
NOTE 2:
Guidance and manufacturer´s declaration – electromagnetic immunity
a)
Field strengths from fi xed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromag-
netic environment due to fi xed RF transmitters, an electromagnetic site survey should
be considered. If the measured fi eld strength in the location in which the device is used
exceeds the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the device.
b)
Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.
Geratherm Medical AG
Fahrenheitstrasse 1
98716 Geschwenda
Germany
www.geratherm.com
IEC 60601
Compli-
– test
ance
level
level
3 Vrms
150 kHz to
3 Vrms
80 MHz
3 V/m
80 MHz to
3 V/m
2.5 GHz
At 80 MHz and 800 MHz, the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and refl ection from structures, objects and people.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the device,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
d= 1.2 √P
d = 1.2 √P
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fi xed RF transmitters, as
determined by an electromagnetic site survey,
be less than the compliance level in each frequency
range.
b
Interference may occur in the
vicinity ofequipment marked
with the following symbol:
34
a
should