3M BV8R Luer Instrucciones De Colocación Y Funcionamiento página 6

Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 35
To reduce the risks associated with chemical exposure, impact, crush, tipping hazards, property damage from
product loss, bypass or cross contamination due to pressure barrier failure, connector failure, purifi cation media
or capsule failure:
• Use only with aqueous based process fl uids in the pH range of 5 to 9.
• Do not use with or expose this product to hot liquids >40°C (104°F).
• Do not use the product on unleveled surfaces.
• Do not use the product for more than 8 hours.
• Do not use this product for continuous service with compressed gasses. Use of compressed gas is permissible for
post-use integrity testing and blow-down purposes only. Limit post-use integrity test and blow-down gas pressures
to less than 1.4 bar (20psig) and temperature less than 25°C for no longer than 30 minutes.
• Do not hang other items on this product.
• Do not gamma or ethylene oxide sterilize the capsules.
• Do not steam-in-place sterilize the capsule.
• Only expose the capsule to 1 sterilization or 1 sanitization cycle.
• Do not autoclave and base sanitize the same capsule prior to use.
• Do not autoclave or base sanitize capsules with the following part numbers BV8 Luer (EMP301AEX020A) and BV8
Sanitary (EMP303AEX020A).
• During pre-use autoclave sterilization do not exceed a temperature of 121°C or a duration longer than 30 minutes for
numbers BV8R Luer (EMP301AEX020R) and BV8R Sanitary (EMP303AEX020R).
• Do not use with or expose this product to process liquids until capsule temperature goes <40°C (104°F).
• Only use NaOH as sanitizing chemical and do not exceed 1M NaOH concentration or 1-hour duration for sanitization.
• Do not pressurize the capsule more than 5 psi during introduction of NaOH solution and static soak during
sanitization.
• Use only 70% ethyl and/or isopropyl alcohol to wipe the exterior of the capsules.
• Do not use with or expose this product to pressurized steam or sanitizing chemicals other than NaOH.
WARNING