6. Malfunctions and Error Messages
Malfunction
Cause
The unit does
The plungers are
not start
either fully advanced
or retracted
Unit starts, but no
Material has set in
paste is extruded
the mixing tip
The plungers are
jammed
The paste flows too
The pastes are too
slowly or not at all
cold, or sealing cap
of the base paste is
defective
The paste is extruded The drive shaft has
but the rotor in the tip not engaged the
does not rotate
mixing tip, it is stuck
in its top position
The cartridge will not
The plungers are not
fit into the unit
fully retracted
The cartridge
receptacle is dirty
The unit does not
The "Start" button
switch off
is stuck
The plungers are
Foil of a bag is
difficult to retract
wedged between
the cartridge and
the plunger
7. Maintenance and Care
7.1 Replacement of the Plunger Discs
If damaged, the plunger discs must be replaced. Only original parts may
be used as replacements to insure safe operation.
Turn off the unit and disconnect the plug from the power source. Use
an Allen wrench to remove the screw in the center of the plunger disc.
Remove the disc and screw a new one into place (Fig. 21).
When changing the plunger discs, make sure you position them with the
correct alignment (see Fig. 21) to prevent damage to the foil bags.
8
7.2 Care
Clean all components with a soft cloth and, if necessary, a mild cleaning
agent (e.g., dish-washing detergent). Solvents or abrasive cleaners may
Remedy
not be used in any case, as they can damage the plastic components!
Turn the plunger
- Cleaning agents must not enter the unit!
adjustment knob to
move the plungers
To disinfect all components, spray the disinfectant on a towel and
from their upper or
use it to disinfect the unit. Do not spray the disinfectant directly on
lower limit
the device.
Attach a new mixing
- Disinfectant agents must not enter the unit!
tip
- Dry residual disinfectants on the device with a soft and fluff-free
Turn the plunger
adjustment knob to
If necessary, ask the manufacturer of the disinfectant if its constant use
check for smooth
will damage plastic surfaces.
movement
Allow the pastes to
7.3 Maintenance and Repair
attain room temperature
(at least 18° C).
No components inside the unit require servicing or repair by the user.
Replace defective foil
3M Deutschland GmbH will accept responsibility for the safety, reliability,
bag
and correct function of the unit only if:
Remove the mixing tip.
- Any modifications or repairs are carried out only by personnel
Turn the plungers
several times up to
- The electrical fittings in the room comply with all regulatory
their upper limit,
thus regaining axial
moveability. Attach
- The unit is operated according to these Operating Instructions.
new mixing tip
Plastic cartridges are subject to wear depending on the material and
Turn the plungers up
frequency of use and are therefore not covered by the warranty for this
to their limit by turning
unit. They should be regularly checked for signs of wear (e.g., fissures)
the adjustment knob
and replaced with new Penta cartridges for Pentamix Lite two years after
clockwise and arresting
initial operation at the very latest. The use of worn or defective cartridges
it. Check the plunger
can have adverse effects on the evenness of the mixture, reducing the
discs for damage
quality of the impression.
Clean the cartridge
receptacle
Press the "Start" button
8. Customer Information
again or unplug the
unit and release the
"Start" button
No person is authorized to provide any information that deviates from the
information provided in this instruction sheet.
Retract plunger
against the resistance,
but avoid damage of
8.1 Warranty
plunger discs
3M Deutschland GmbH warrants this product to be free from defects
in material and manufacture. 3M Deutschland GmbH MAKES NO
OTHER WARRANTIES, INCLUDING ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. User is
responsible for determining the suitability of the product for user's
application. If this product is defective within the warranty period, your
exclusive remedy and 3M Deutschland GmbH's sole obligation shall be
repair or replacement of the 3M Deutschland GmbH product.
8.2 Limitation of Liability
Except where prohibited by law, 3M Deutschland GmbH will not be liable
for any loss or damage arising from this product, whether direct, indirect,
special, incidental or consequential, regardless of the theory asserted,
including warranty, contract, negligence or strict liability.
cloth, as they damage the plastic components.
authorized to do so by 3M Deutschland GmbH;
requirements;