Note!
• While you adjust the values, the car-
riage will move to the set range. This
allows you to easily and quickly de-
termine the ROM where the patient
does not experience any pain.
• The angle entered last for the respec-
tive directions of motion will be saved
in each case.
5.3 Therapy parameter details
• To access the programming level you
need, repeatedly press the MENU key.
• The treatment parameters are selected
with the different parameter keys.
• To change the treatment values, press
the + / – keys.
• To enable/disable a function, press the
respective parameter key again.
• To save the entered values, press the
STOP key.
LEVEL 1:
n extension
maximum value:
n flexion
maximum value:
n pronation
maximum value:
n supination
maximum value:
Note!
• The programmed values and the
values (degrees) measured on the
patient may deviate slightly.
• To ensure a physiological motion, the
carriage simultaneously moves to the
following positions in the synchro-
nized mode:
− maximum extension value and
maximum pronation value
− maximum flexion value and
maximum supination value.
- 5 degrees
140 degrees
n Pauses
- 90 degrees
90 degrees
67
• To ensure that the movements you
program are physiological in nature,
we recommend to program the car-
riage in the following order:
1. maximum extension value
2. maximum pronation value
3. supination
(to relieve the elbow)
4. maximum flexion value
5. maximum supination value
• When the START key is pressed in the
synchronized mode, the carriage will
first move to the position halfway bet-
ween the set maximum extension/
flexion values as well as the prona-
tion/supination values. The carriage
will stop in this position. Treatment
begins when you press the START
key again.
• Horizontal adduction/abduction is
adjusted manually. These are possible
settings:
horizontal adduction :
horizontal abduction:
LEVEL 2:
Pauses occur at each programmed
maximum value.
These are the two pause points:
− maximum extension value and
maximum pronation value
− maximum flexion value and
maximum supination value.
Pauses are separately adjustable for mo-
tor A and motor B between 1 second and
60 minutes. Pauses can be set in steps of
1 second to any value between 0 and 59
seconds, and in steps of 1 minute to any
value between 1 and 60 minutes.
Default: no pause
= 0°
120°
0°