62
3.3 Function specifications
3.3 Function specifications
underline: default value
LCD display
Display refresh
rate
Data hold function
Auto-power-save
(APS) function
Backlight
function
Alarm function
Resistance
alarm function
Current alarm
function
Alarm Hi/Lo
Alarm threshold
setting range
Filter function
Resistance
measurement
filter function
Current
measurement
filter function
Memory function
Memory
capacity
Up to 2000 counts
500 ms (approx. 2 times/s)
Instrument automatically turns off approx. 5 min. after last key
operation.
Automatically turns off approx. 2 min. after last key operation.
Resistance measurement mode alarm: Beeps when measured
value is less than or greater than threshold.
Current measurement mode alarm: Beeps when measured value
is less than or greater than threshold.
Separate Hi/Lo settings for resistance measurement and current
measurement
Resistance measurement: Hi.AL/Lo.AL
Current measurement: Hi.AL/Lo.AL
Resistance measurement: 0.02 Ω to 1600 Ω
Resistance measurement initial value: 25.0 Ω
Current measurement: 0.05 mA to 200.0 mA, 0.201 A to 60.0 A
Current measurement initial value: 1.00 mA
Moving average time: Max. 9 sec.
Cutoff frequency: 180 Hz ±30 Hz (-3 dB)
2000 values
HIOKI FT6380B983-00