The skin becomes
The electrodes are
red and/or you feel
dirty.
a stabbing pain.
The surface of
the electrode is
scratched.
Output current
The electrodes
stops during
come off the skin.
therapy.
The power of the
batteries has been
exhausted.
The Remote control
Are the batteries
can not power on
installed correctly?
The Remote control
The
can't connect to the
connection failure.
EM 70 device and
the remote control
displays E1.
13. Technical Specifications
Model:
Type:
Suggested charger:
Main unit Lithium Ion
Battery:
Clean the elec-
trodes according to
description in this
manual or replace
new electrodes.
Replace new
electrode.
Turn off the device
and place the
electrodes again.
Charging.
Insert the batteries
observing polarity.
Restart the EM 70
Bluetooth
®
and the Remote
control.
EM 70
LT1102S (Unit) /
LT1102S-R (Remote controller)
I/P: 100-240V~, 50/60Hz, 0.2A.
O/P: 5V , 300mA.
3.7V
, 250mAh, 0.925Wh
Remote control:
Frequency:
Pulse Width:
Wave form:
Output Voltage:
Max current:
Output Intensity Level:
Operating Conditions:
Storage and transportation
conditions:
Size: Main unit:
Remote controller:
Weight: Main unit:
Remote controller:
Service life of the device:
Service life of electrode
pads:
Applied part:
IP classification system:
Classification of ME
Equipment:
Maximum separation
distance:
The recommended
separation distance:
36
I/P: 2×1.5V AAA batteries
2Hz~290Hz
30us~300us
Biphasic square wave
max. 45V (at 1000 ohm load)
45mA (at 1000 Ohm)
0~15 levels
5 °C~40°C;
30%RH~75%RH; 700hPa ~1060hPa
14°F~131°F (-10°C~55°C);
10%RH~90%RH; 700hPa ~1060hPa
72(L)×50(W)×14(H)mm
127(L) ×48.6(W) ×24.4(H)mm
approx. 38 g
approx. 78 g(with batteries)
2 years
10~20 times
Electrode
IP22
Internally powered(operating) / Class
II(charging)/ Continuous operation
10m (In the opened environment)
3m