Choose the desired selection by pressing the CYCLE
push-button.
F R O N T
/
R E M O T E
S E L E C T
I O N :
•
Press PUMP CURRENT push-button and the display
shows:
S E R I
A L
This message will only be displayed if the RS 485 module
is installed.
Where:
XX = 00 ÷ 31 is the unit address
/
S E R
F R O N T
A D D R E S S
X X
•
Press the PUMP CURRENT push-button to enter the
value and ,if the serial option is installed, the display
shows:
R S X X X
B A U D
R A T E
where: BAUD RATE XXXX = means 600, 1200, 2400,
4800, 9600 baud rate for the host computer or printer
communication.
The controller is factory-set for 9600 baud rate operation.
The value which corresponds to the board installed will be
assigned to RSXXX.
•
If the RS 232 is installed, enter the value by pressing
the PUMP CURRENT push-button and, if the serial
option is installed, the display shows:
R S X X X
S E L E C T
This message will only be displayed if the RS 232 interface
is present.
where:SELECTION: XXXXX = means HOST or PRINT.
The controller is factory-set to HOST
84
TECHNICAL INFORMATION
X X X X
H O S T
/
P R I
I O N :
X X X X X
87-900-904-01 (I)
N T