a Sipper module NANOCOLOR
b Peristaltic pump
c Supply hose
d Drain hose
e Flow cuvette
10. Operation and menus
10.1 Activating the sipper module on the spectrophotometer
After connecting the sipper module to the spectrophotometer, open the connectivity menu
via
→
→
confirm the changes by pressing
spectrophotometer. The
10.2 Setting the pumping, rinsing, and settling times
The sipper module settings are configured in a separate settings menu on the spectropho-
tometer. After connecting the sipper module to the spectrophotometer, open the configuration
menu via
→
A window displaying the properties of the sipper module appears.
Pumping time:
28
®
FP-200
. Activate the checkbox for the FP-200 and the USB A interface. Please
. The sipper module is detected automatically by the
symbol appears in the status bar of the photometer.
.
→
The pumping time determines the sample volume that will be
pumped through the flow cuvette. To ensure a 50 mm flow cuvette
is completely rinsed, the pumping time must be set to a min. of 10 s.
NANOCOLOR
f Connecting hose
g Photometer
h USB connecting cable
i Hose adapter
®