10.7 Performing a scan

Open the menu for scans via
Hold the intake hose in the zero solution and press the
into the flow cuvette and the zero measurement is taken automatically after the settling time
has expired. The time remaining until the measurement will be taken is shown in the display.
You can cancel the measurement by pressing
Hold the intake hose in the sample solution to be analyzed and press the
sample solution is pumped into the flow cuvette and the sample measurement is taken
automatically after the settling time has expired. The time remaining until the measurement
will be taken is shown in the display. You can cancel the measurement by pressing
The flow cuvette is rinsed using distilled water or the next sample solution. The rinsing oper-
ation is initiated in the corresponding measurement menu by pressing
complete, the results are displayed as a curve in the result window.
You can enter various information on the sample using the icons below the result window.
The information entered is saved when you exit the menu or when the next measure-
ment is taken. The scan can be edited using the scan analysis and the
®
NANOCOLOR
VIS II and
34
. Enter the required information and confirm with
/
II spectrophotometer manual).
UV
VIS
NANOCOLOR
icon. The zero solution is pumped
.
®
FP-200 V 1.00 / 11.17
.
icon. The
.
. After the scan is
icon (see the