can indicate wrong functioning. A protection fuses (F 1 x 1.6 A L 250V) reachable from exterior and situa-
ted in the plug protects the instrument.
For fuses replacing, always the type and the range. Before changing the fuse, disconnect the plug from
the power supply socket.
Internally, the device (only for devices fitted with a circuit board) is protected by a fuse (T 50mA L 250V)
that cannot be reached from the outside, so please contact a technician authorised by the manufacturer
for its replacement.
Fault type
1. The suction unit doesn't
work
2. No aspiration
3. No aspiration
4. The Vacuum power on the
patient side is either very low
or absent
5. The float doesn't close
6. The float doesn't close
7. Low suction
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8
If the overfill security system it's activated, don't proceede with the liquid aspiration.
If the overfill security system doesn't work there are two cases:
1° case – If the overfill security system doesn't work the aspiration will be stopped by the bacteriological
filter who avoid the liquid penetration inside the device.
2° case – If both the security system doesn't work, there is the possibility that liquid comes inside the
device, in this case return the device to GIMA technical service.
GIMA Srl will provide upon request electric diagrams, components list, descriptions, setting instructions
and any other information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA TECHNICAL SERVICE.
GIMA DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING,
APPEARS TO BE TAMPERED.
INSTRUCTIONS
The working position must be such as to allow one to reach the control panel and to have a good view
of the empty indicator, the jar and the antibacterial filter.
If the device is to be transported from one place to another, to prevent the liquid collection jar from
falling and consequently the liquid from spilling, removing the jar from the device is recommended.
eNgLIsh
Cause
Cable is damaged
External power source failure
Jar Cap badly screwed down
Lid seal not in its seat
a) Vacuum regulator set to minimum
b) Protection filter blocked or dama-
ged
c) Connection tubes blocked, kinked
or disconnected
d) Shut-off valve blocked or dama-
ged
e) Pump motor damaged
If the cap has been washed, ensure
that the float is not partially detached
The float it's covered by dirty mate-
rial
Foam inside the jar
None of the remedies has achieved
the desired results
Solution
Replace the cable
Check the external power source
Unscrewed the cap, then rescrew it
correctly
Unscrew the cap and insert the seal
properly in its seat
a) Turn the vacuum regulator clockwise
and check the value of the vacuum on the
gauge
b) Replace the filter
c) Replace or reconnect the tubes, check
the jar connections
d) Empty the jar, or disconnect the tube
from the jar and unblock the shut-off val-
ve. The unit twill only work in the upright
position
e) Refer to authorised service personnel
Insert the float into it's place
Unscrewed the cap, leave the and put in
on autoclave
Fill the jar to 1/3 full of ordinary water
Contact the seller or GIMA After-sales
Assistance Service