WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full volume
of use of the jar and better efficiency of the overflow device. The vacuum jar, during use, must be used
in vertical mode, to prevent the action of the backflow valve. If this protection is triggered, turn the device
off and disconnect the pipe connected to the vacuum jar (indicated with the word VACUUM) on its cover.
WARNING: For proper use, place the aspirator on a flat, stable surface in order to have the full
volume of use of the jar and better efficiency of the overflow device. The vacuum jar, during use, must
be used in vertical mode, to prevent the action of the backflow valve. If this protection is triggered, turn
the device off and disconnect the pipe connected to the vacuum jar (indicated with the word VACUUM)
on its cover.
WARNING: Ensure that the IN marker on the filter is on the side facing the collection jar lid and fitted
into the "VACUUM".
A wrong connection causes immediate destruction in case of contact with sucked liquids.
Suction Pump Inlet
Connect the short silicon tube 1, with antibacterial
filter 2, to the suction connector 3. The other
tube, with one end connected to the filter must be
connected with the other end to the jar's connector
4 where has been fixed the red float (security float).
When the 90% of the volume of the jar is reached
there is the activation of the security float (the float
close the aspiration connector on the jar) to avoid
liquid penetration inside the device. The device
must be used on a plan of horizontal operation.
Connect the long silicon tube 5 to the other jar's
lid connector 6
Connect the other end of the long silicon tube 5 to
the probe plastic connector 7
Connect the power cord to the device than connect
the plug to the electrical mains supply.
Unscrew the jar's lid and fill the jar 1/3 full or
ordinary water (this for an easy cleaning operations
and an rapid reaching of the functionality vacuum)
then rescrew the lid on the jar correctly.
Push switch 8 on position I to start suction
Once finished push switch on O position and
unplug
To extract the accessories and start with cleaning it.
WARNING: The power supply cable plug is the element of separation from the electrical mains system:
even if the units equipped with a special on / off switch button, the power supply plug must be kept ac-
cessible once the device is in use so as to allow a further method of disconnection from the mains supply
system.
FLOW DIRECTION
eNgLIsh
Jar Air Tube