Binder C 150 Instrucciones Resumidas página 5

Incubadora de co2
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

Quick Guide
Do NOT introduce any combustible or explosive materials into this
!
device. Prevent the release of combustible mixtures of solvents
with air and dust.
Some parts of the device may become hot during the hot-air
sterilisation - Do NOT touch.
This quick guide does not replace the operating manual. Read the
operating manual carefully before taking this unit into operation and
familiarize yourself with all safety relevant information.
Controller RP1
Main power switch
handle
Door
Lower
housing cover
BINDER GmbH
Postfach 102
78502 Tuttlingen
Service e-mail: service@binder-world.com
GB
GB
CO Incubator C 150 (E2)
GB
2
Tel.: +49 (0) 7462 2005 0
Fax: +49 (0) 7462 2005 100
http://www.binder-world.com
E-mail: info@binder-world.com
Service hotline: +49 (0) 7462 2005 555
Service fax: +49 (0) 74 62 2005 93 555
Rear connection panel
11
12
(11)
Power cable
(12)
Miniature fuse
(13)
Connection socket for zero-voltage relay alarm contact
(14)
Quick acting closure socket for CO gas bottle
(15)
Ethernet interface (option)
Connection and putting into operation
Please check the power supply before connecting the unit and before the first run. Compare the values
with the data given on the type plate (left side, lower front).
Connect the device to a power supply of the specified ratings.
Set the gas supply pressure of the gas bottle to 2.0 bar above ambient pressure. Connect the gas bottle at
the rear of the unit to the corresponding socket. Connection for hose DN 6. Use CO gas with a technical
grade of 99.5 percent.
The gas connections must be established by qualified personnel who are trained in
handling
CO gas
and familiar with the required safety measures.
2
relevant regulations for handling gas cylinders and
!
Do NOT exceed the indicated output pressure of 2.5 bar above the ambient
pressure. Before connection, check the output pressure at the pressure reducer of
the gas supply.
If several units with different maximum values of gas admission pressure are connected to a common
supply net, use pressure reducers for those units with a lower maximum gas admission pressure.
Plug in the CO sensor without turning into the intended connection bushing in the upper part of the rear of
2
the inner chamber. Fill the water pan with distilled, sterilized water.
Copyright
©
10/2012
BINDER GmbH
13
14
15
2
2
CO gas
.
2
Art. No. 7001-0235
Observe the

Publicidad

Tabla de contenido
loading

Tabla de contenido