interpreted as qualitative only. As with any diagnostic procedure, the test results indicated on the
screen are interpreted in the light of the patient's clinical presentation and any other information
available to the physician.
M
AINTENANCE AND CLEANING OF THE
The Reader is maintenance free.
Before each test, the glass on the bottom (covered by the slide holder) should be checked for
impurities. For cleaning a commercial cloth together with a commercial cleaning fluid, e.g. for
glasses products, is recommended.
Due to the possibility of contact with potentially infectious material while using the Reader,
disinfection is required to prevent infections or other contaminations. For disinfection every surface
of the Reader as well as the slide holder needs cleaning with a suitable disinfection solution. That
solution approved for medical devices, should not affect the device housing material. The
disinfection spray Mikrozid® AF Liquid for instance is a suitable solution. Comparable products can
be used as well.
For cleaning the display use dry cloth. Never apply fluid onto the display!
S
,
TORAGE
STABILITY AND TRANSPORTATION TERMS
Store the slide and the Control sample
in the manufacturer's packaging;
in a dark, dry place with the temperature from +15 °С to +50 °С;
in a place protected from mechanical actions (friction, pressure, strokes);
keep away from the ammonia vapor, moisture and direct sunlight.