Do not clean, resterilize, or try to extract the needle from the needle housing. An accidental needle stick may
occur.
Do not reuse sensors. Reuse of the sensor may cause damage to the sensor surface and lead to inaccurate
glucose values, site irritation, or infection.
Where to insert the sensor
CAUTION: Avoid the 2 inch (5.0 cm) around the navel to help ensure a comfortable insertion site, and
to help with adhesion.
Choose an insertion site that has an adequate amount of subcutaneous fat. The Enlite sensor has been
approved for use on the abdomen. See step 2 for a diagram of these areas.
Do not insert the sensor in muscle or areas constrained by clothing or accessories; areas with tough skin or
scar tissue; sites subjected to rigorous movement during exercise; or in sites on the belt or waistline to avoid
chafing or site irritation.
Storage and handling
Store the Serter in the released position, at room temperature, to maintain optimum product performance and
life.
Specifications
Dimensions
2.93 x 1.95 x 2.45 inches
Serter life of use
The Serter can be used for up to 600 sensor insertion uses. And, it has been tested through 92 disinfection
cycles for functionality.
WARNING: The disinfection process may damage the Serter prior to 600 uses. Always inspect the
Serter for any signs of cracking, flaking, or damage prior to use. Discard the Serter if you see any of
these signs to avoid incorrect sensor insertion, pain or injury.
Serter maintenance
The Serter is intended for single-patient use. Always clean and disinfect the Serter after each use.
To clean and disinfect the Serter, you will need to following materials: gloves, mild liquid soap (for example
Ivory
liquid soap), household bleach (for example Clorox
®
and a container.
To clean the Serter:
1. Prepare a mild liquid soap solution using one (1) teaspoon of mild liquid soap per one (1) gallon of tap
water.
2. Rinse the Serter under room temperature tap water for at least one minute. Continue rinsing until the
Serter is visibly clean. Make sure all hard-to-reach areas are rinsed completely.
3. While rinsing the underside of the Serter, press and release the mechanism on the underside of the
Serter to ensure that the underside of the Serter is completely rinsed with water.
4. Submerge the Serter in the mild liquid soap solution and soak for at least 10 minutes.
5. Brush the entire surface of the Serter using a soft-bristled toddler toothbrush until visibly clean. Pay close
attention to brushing the hard-to-reach areas.
6. Rinse the Serter under running room temperature tap water for at least one minute, and until all visible
liquid soap is gone.
7. Shake off any excess water or moisture, and place the Serter upright on a clean, dry, cloth to air-dry.
To disinfect the Serter, continue with steps 8 through 11:
Weight
71.2 grams
Regular-Bleach), soft-bristled toddler toothbrush,
®