INSTRUCTION MANUAL CODE 80243 REV A
Figure 18
Figure 19
J
.P. SELECTA s.a.
January/11
Press [RETURN] to go back to pre-testing interface.
Note:
If you want to delete an equation, move the cursor onto it, and just press [CLEAR].
Test and Print
When you have selected the equation, press [START/STOP] to go into testing
interface in pre-testing interface. (Fig. 19)
After blanking, pull the unknown sample cuvette in the light path, then press [EN-
TER]. The testing result will be displayed on the screen.
Operation is the same if you want to test other samples.
Press [PRINT] to print testing results. (Fig. 20)