Lamp Life Management; Load Default; Lamp Change; Version Of Instrument - J.P. SELECTA UV2005 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80243 REV A
Figure 49
Figure 50
Figure 51
J
.P. SELECTA s.a.
January/11
lamp. (Fig. 49)
When the used time of D2 lamp is higher than 1000 hours, replacing a new one is
strongly recommended.
When the used time of W lamp is higher than 1000 hours, replacing a new lamp is
strongly recommended.
When you have replaced a new lamp, you'll have to go into the lamp life
setting screen, and move the cursor on the "Reset W Life" or "Reset D2 Life" by
pressing [»] and [«] followed by [ENTER] to reset them into zero.

7.8 Load Default

Move the cursor onto Load Default by pressing [»] and [«] in utility setting screen,
and then press [ENTER]. (Fig. 50)
If you want to default the parameters, choose "Yes" in the lower screen of Fig. 50.
If not, you should choose "No".
Press [RETURN] to exit without any choice.

7.9 Lamp Change

Move the cursor on "Lamp Change" by pressing [»] and [«] in utility setting screen,
then press [ENTER]. (Fig. 51)
Input the new wavelength you want by pressing the numeric keypad followed by
[ENTER].
Lamp change wavelength means the point at which W lamp and D2 lamp exchange.
The range of it is 300 - 400, other value is invalid.

7.10 Version of Instrument

Move the cursor onto Version by pressing [»] and [«] in the utility setting screen,
and then press [ENTER]. Then you will find the detailed messages of the instrument.
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(It can be modified without notice)
Tel 34 93 770 08 77 Fax 34 93 770 23 62
Page: 51