MANUAL DE INSTRUCCIONES CODIGO 80243 REV A
Figure 33
Figure 34
Figure 35
J
.P. SELECTA s.a.
Enero /11
Move the cursor onto Total Time by pressing [»] and [«], followed by [ENTER].
(Fig.33)
Input the new total test time by pressing the numeric keypad followed by [ENTER],
and it will return to the kinetics setting screen.
Note:
1. The range of total test time is 1-120,000, other value is invalid.
2. The maximum value of total time is correlative to the interval time, because system
only permits that 1000 dots will be adopted at most.
For example:
if the interval time is 1s, then the maximum total test time is:
if the interval time is 20s, the maximum total test time is:
if the interval time is 200s, the maximum total test time is NOT:
However, due to the reason that the range of total time is 1-120,000, when the interval
time is 200s, the maximum test time will be 120,000s!
3. If you don't want to change the value, press [RETURN] directly.
Set Test Mode
Move the cursor onto Test Mode by pressing [»] and [«] in the setting screen,
followed by [ENTER]. (Fig. 34)
Move the cursor onto the test mode by pressing [»] and [«], and then press [ENTER]
to select. Press [RETURN] to go back to the kinetics setting screen.
Set Upper Limit
Move the cursor onto Upper Limit by pressing [»] and [«] in the setting screen,
and then press [ENTER]. (Fig. 35)
Input the new upper value by pressing the numeric keypad followed by [ENTER], it
will return to the kinetics setting screen.
If you don't want to change the value being set last time, press [ENTER] or [RE-
TURN] directly.
Note:
The range or A is -0.3 - 3, the range or T is 0 - 200, other value is invalid.
Set Lower Limit
Move the cursor onto Lower Limit by pressing [»] and [«] in the setting screen,
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
(Sujetas a modificaciones sin previo aviso) Pag.: 46
1 X 1000=1000s;
20 X 1000=20,000s;
200 X 1000=200,000s
Tel 34 93 770 08 77 Fax 34 93 770 23 62