Troubleshooting and Error message
Error
Problem
Message
ERR0
Exceed the zero range
Initial zero point exceeds 30% of
ERR2
full load
ERR3
Exceed the A/D resolution range
ERR4
EEPROM failure
ERR5
Overload condition
ERR6
Exceeds the display range
Accumulated number of
ERR7
weighments exceeds the display
range
Weight limit value is higher than the
ERR8
full load value
ERR9
Exceed tare or pre-tare range
ERR10
Wrong calibration weights
Exporting data to PC in the form of EXCEL
Introduction:
Connect the scale with PC and set the parameter of external device as "EXCEL" on the
scale, then you could export the weighing data to PC in the form of EXCEL. With this
function, you could record/accumulate/average/data statistical analysis the testing data,
which we could call it as scale-computer data management function.
Note: pls enable "Use Serial Keys" function in the computer.
Solution
The item on the pan should be within 2% of
full load.
1.Remove the obstacle article under the pan
2.Replace the load cell or contact the
maintenance department.
1. Replace A/D
2..Replace the load cell or contact the
maintenance department.
Re-weld
EEPROM
maintenance department.
Remove the overload item
-----------------------------------------------
Delete the exceeding weights
Reset the weight limit value.
0<Tare value≤ full load
Place the right test weights and the
calibration value should be below full load.